Skip to content

Public FDA record

Z-2206-2015: American Science & Engineering Inc,Gemini Cabinet X-ray system.

Recall record Z-2206-2015 for American Science & Engineering, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2206-2015
Company in FDA record
American Science & Engineering, Inc.
FDA classification date
Aug 14, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

During factory testing, we discovered that Gemini systems could, in rare circumstances, experience a fault that leads to a state in which X-rays are being produced but the X-ray indicator lights (located on the operator's console and on both ends of the tunnel) are not illuminated.

Source facts

Action
American Science Engineering, Inc (AS&E) is developing a fix for this problem that includes updating a software configuration on the X-ray source. You will be notified by field service as soon as this update is available. Please inform all operators of the system about this communication, so they are aware of the need to shut down and re-start the system should this fault occur. We appreciate your understanding. if you have any questions, please feel free to contact AS&E via e-mail at service@as-e.com. If you are under a service contract, you can also contact your customer service representative. For further questions please call (978) 262-8700.
Initiated
2015-05-22
Root cause
Vendor change control
Terminated
2017-02-22
Distribution
Worldwide Distribution - US Distribution to the states of : NY, FL, TN, PA, NJ, DE, ME, TX, CT, MS, NC, MD, MO, VA, DC, NV and GA., and to the countries of : Saudi Arabia, Indonesia, Canada, Netherlands, Hong Kong, Lebanon and Kazakhstan.
Recall number
Z-2206-2015
Recall status
Terminated
CDRH record ID
138765
FDA classified
2015-08-14
Recall event number
71696

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗