Skip to content

Public FDA record

Z-2206-2020: Medtronic Implantable Insulin Pump System, REF: MMT-2007D, Sterile EO, CE 0459 UDI: 00763000043001

Recall record Z-2206-2020 for Medtronic Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2206-2020
Company in FDA record
Medtronic Inc.
FDA classification date
Jun 3, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Internal testing results showed endotoxin-related test result did not meet specifications for implantable insulin pump. If pumps have higher than allowed endotoxin levels, patients may have an effect of fever, hypotension, anaphylactic shock, hyperglycemia, or diabetic ketoacidosis (DKA), potentially followed by secondary multi-organ failure (primarily renal and hepatic) and/or death.

Source facts

Action
Medtronic initiated verbal communication with affected consignees on April 10, 2020 via phone call and provided affected consignees with an "Urgent Field Safety Notice" via email between the dates of 04/24/20 - 4/30/20. . In addition to informing consignees about the recall, the firm asked consignees to take the following actions: 1. Immediately identify and quarantine all unused affected pumps in your inventory. 2. Return all unused affected pumps in your inventory to Medtronic. Your Medtronic Representative can assist you in the return and replacement of this pump as necessary. 3. Forward this notice to all those who need to be aware within your organization and to any organization where the affected pumps may have been transferred. Please maintain a copy of this notice in your records.
Initiated
2020-04-10
Root cause
Process control
Terminated
2021-04-27
Distribution
No US Distribution. OUS: Netherlands, France, and Sweden
Recall number
Z-2206-2020
Recall status
Terminated
CDRH record ID
181272
FDA classified
2020-06-03
Recall event number
85582

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.