Skip to content

Public FDA record

Z-2206-2023: Medline O.R. Scissors, S/B, Sterile, 5.5, Model #DYND04000

Recall record Z-2206-2023 for MEDLINE INDUSTRIES, LP - Northfield. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2206-2023
Company in FDA record
MEDLINE INDUSTRIES, LP - Northfield
FDA classification date
Jul 20, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The scissors were manufactured without a tip protector resulting in the scissors breaking through the semi-rigid plastic tray compromising sterility.

Source facts

Action
The recalling firm issued letters dated 5/12/2023 via first class mail and email on 5/12/2023. The letter explained the reason for recall and provided the required actions. The actions were to immediately check their inventory for the affected item number and lot numbers listed on the recalling firm's recall portal and destroy the products. A website link is provided for completing the response form and for obtaining the specific item and lot numbers affected. Once the response form is submitted, the consignee account will be credited. Distributors are instructed to notify the customers of the recall who have received the affected product and the customer is to document the destruction of their product. The distributor is to include the customer product quantities in their response form.
Initiated
2023-05-12
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution. There was government distribution but no military or foreign distribution.
Recall number
Z-2206-2023
Recall status
Open, Classified
CDRH record ID
200872
FDA classified
2023-07-20
Recall event number
92412

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗