During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior
has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0,
an actual x-ray dose is applied to the patient.
Source facts
Action
Distribution of Customer Advisory Notice (CAN) to all affected customers began by Siemens
on March 9, 2026.
Initiated
2026-03-09
Root cause
Radiation Control for Health and Safety Act
Distribution
US and Worldwide
Recall number
Z-2206-2026
Recall status
Open, Classified
CDRH record ID
220380
FDA classified
2026-06-04
Recall event number
98992