Skip to content

Public FDA record

Z-2207-2020: NobelActive Internal NP 3.5x15mm Endosseus dental implant, REF: 34128, Rx Only, Sterile R, CE, UDI: (01)0

Recall record Z-2207-2020 for Nobel Biocare Usa Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2207-2020
Company in FDA record
Nobel Biocare Usa Llc
FDA record creation date
Jun 3, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The top label for dental implant may incorrectly indicate the implant length. Specifically, the label may contain 11.5mm for the length, which is shorter than the actual length of the implant (15mm). An incorrect sized implant may cause patient discomfort, permanent nerve damage and may require an additional surgery.

Source facts

Action
Customers who received product/affected lot were to be contacted via Urgent Medical Device Recall Letter dated July 18, 2019 via USPS on July 19, 2019. In addition to informing customers about the recall, the recall notification asked customer to take the following actions: 1. Identify affected devices in your stock by confirming the information on the side label: NobelActive Internal NP 3.5x15mm, article 34128, lot 13080792. 2. Confirm the information on the top label as shown in figure 1 above. 3. Complete attached Customer Acknowledgment Form and return it to Nobel Biocare, via email to us.complaints@nobelbiocare.com or fax 714-282-5073, within 5 days of receipt of this letter. 4. Return incorrectly labeled products (top label 3.5x11.5) to: Nobel Biocare Attn: Returns PFA1917 22715 Savi Ranch Pkwy Yorba Linda, CA 92887 . 5. Correctly labeled products (top label 3.5x15) do not need to be returned. 6. If you require any further information or support, please contact your customer support representative at 1-800-322- 5001 x.1527.
Initiated
2019-07-19
Root cause
Labeling mix-ups
Terminated
2020-10-27
Distribution
US: AL, AR, AZ, CA, CO, CT, FL, HI, ID, IL, IN, KY, MA, MS, MI, MO, NC, NH, NV, NY, OK, PA, SC, SD, TX, UT, WA, WI OUS: None
Recall number
Z-2207-2020
Recall status
Terminated
CDRH record ID
180559
FDA record created
2020-06-03
Recall event number
85331

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗