Z-2208-2010: Access Immunoassay Systems AFP QC, Part Number: 33219 The Access AFP QC is intended for monitoring syste

Recall record Z-2208-2010 for Beckman Coulter Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2208-2010
Company in FDA record
Beckman Coulter Inc.
FDA classification date
Aug 11, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The recall was initiated after Beckman Coulter confirmed an issue with the Access AFP QC kits (REF 33219) which contain three levels of control material identified as QC1, QC2, and QC3, Beckman Coulter has confirmed that vials of QC1 in the kit lots identified above contain microbial contamination. Customers may see that the results from a contaminated vial of QC1 are within the established refere

Source facts

Action
A Product Corrective Action (PCA) letter was sent on the week of April 19, 2010 to all customers who purchased the Access Immunoassay Systems AFP QC material, lot numbers referenced above. The letter provided the customers with an explanation of the problem identified and instructed the customers to (1) Please discard all remaining inventory of the three Access AFP QC kit lots identified above, including all vials of QC1, QC2, and QC3. (2) Please review your ac and patient results obtained while using the affected control lots. Interpret Access AFP test results in light of the clinical presentation of the patient, including: symptoms, clinical history, data from other tests and other appropriate information. (3) Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. (4) Complete and return the enclosed response form within 10 days so we can be assured that you have received this important notification. For customers that need assistance or have any questions regarding this notification, they were instructed to contact Technical Support at 1-800-854-3633 in the United States and Canada.
Initiated
2010-04-21
Root cause
Other
Terminated
2012-05-07
Distribution
Nationwide and Canada
Recall number
Z-2208-2010
Recall status
Terminated
CDRH record ID
91542
FDA classified
2010-08-11
Recall event number
55678

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗