Open original FDA record ↗Recall record
- FDA identifier
- Z-2208-2023
- Company in FDA record
- Baxter Healthcare Corporation
- FDA classification date
- Jul 20, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
An issue was identified where the LCD monitor display may not indicate the correct active channel. The issue occurs when the monitor is turned on using battery power, and channel B was selected as the last active channel prior to powering down.
Source facts
- Action
- Synovis Micro Companies Alliance (MCA), Inc., a subsidiary of Baxter International Inc., is issuing an Urgent Medical Device Correction to its consignees on 06/19/2023 by USPS First Class Mail. The notice explained the problem with the device, risk, and provided the following method to circumvent the problem: "The affected flow coupler monitors may continue to be used. To avoid encountering this issue, please use the monitor while connected to an AC power supply whenever possible. However, if using battery power, Baxter advises users to select channel A prior to shutting down the monitor. Upon powering the monitor back up, please check to ensure that the channel illuminated via the hardware button matches the LCD screen channel display. If a mismatch occurs, touch the channel display on the LCD screen or press the hardware button for channel B to resolve the issue. For general questions regarding this communication, please contact Synovis MCA at 800-510-3318 or 205-941-0111, between the hours of 8:00 am and 5:00 pm Central Time, Monday through Friday."
- Initiated
- 2023-06-19
- Root cause
- Under Investigation by firm
- Distribution
- US
- Recall number
- Z-2208-2023
- Recall status
- Open, Classified
- CDRH record ID
- 201295
- FDA classified
- 2023-07-20
- Recall event number
- 92525