Skip to content

Public FDA record

Z-2209-2008: VectorVision (VV) Sky Navigation Platform (19" Computer Rack); the 19" computer rack is a component of th

Recall record Z-2209-2008 for Brainlab AG. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2209-2008
Company in FDA record
Brainlab AG
FDA classification date
Sep 17, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Diameter of cables used for installation are to small for the applied current. If an internal short circuit is produced the medical power supply will not shut down automatically and will continue to deliver current, which could result in overheating cables.

Source facts

Action
BrainLAB sent Product Notification letters dated 11/12/07 to their VectorVision Sky end users, informing them of the possible overload of electrical cables in VVsky products with Data Billboards and/or microscope integrations and/or Device Connections Panels (Wall Panels). The users were requested to perform an inspection of the medical power supplies and cabling on a regular basis. If anything suspicious is noted, they were directed to shut down the entire system and contact BrainLAB immediately. BrainLAB contacted each account shortly after the sending of the letter to arrange for the installation of small fuses/circuit breakers according to the specific load limits per device, which will open the circuit in case of overload on faulty circuits, and to exchange the low voltage cables that directly connect the medical power supply from 1.5 mm2 to 2.5 mm2 thickness to withstand the full 20 amp current in case of a failure. Any questions were directed to BrainLAB's customer hotline at 1-800-597-5911.
Initiated
2007-11-12
Root cause
Component design/selection
Terminated
2009-07-24
Distribution
Nationwide Distribution --- including states of Arizona, Connecticut t, Idaho, Massachusetts, Michigan, Minnesota, New York, Ohio, Pennsylvania, Texas, Virginia and Wisconsin.
Recall number
Z-2209-2008
Recall status
Terminated
CDRH record ID
71856
FDA classified
2008-09-17
Recall event number
48682

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗