Certain lots of COBAS AmpliPrep (CAP) S-tube Input (Box of 12 bags 12x24 S-tube and 12x24 S-clip) can cause the sample input tubes to stick in the barcode clips. As a consequence, the CAP instrument will lift both the sample input tube together with the barcode clip from the sample rack, resulting in an aborted run. No erroneous results would be produced if the issue occurs, but samples in proce
Source facts
Action
On 1/17/2012 an "Analyzer Bulletin" notification was sent to Roche affiliates and on 2/1/2012 an "Analyzer Bulletin" notification was sent to all affected US Consignees. The notification described the product issue and provided recommended actions. Customers were referred to Roche Technical Support at 800-526-1247 for questions.
Initiated
2012-02-01
Root cause
Nonconforming Material/Component
Terminated
2013-07-16
Distribution
Nationwide Distribution (USA).
Recall number
Z-2209-2012
Recall status
Terminated
CDRH record ID
111094
FDA classified
2012-08-15
Recall event number
62586