Open original FDA record ↗Recall record
- FDA identifier
- Z-2209-2014
- Company in FDA record
- Medical Positioning Inc
- FDA classification date
- Aug 13, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Excessive weight on the table will cause tables with the affected component to drop into the Trendelenburg position more rapidly than normal. The product was manufactured with a component that can cause the device to malfunction.
Source facts
- Action
- Medical Positioning Inc. sent Potential Product Hazard Bulletin letter dated July 8, 2014 to affected customers, via USPS certified return receipt mail. The letter identified the affected product, problem and actions to be taken. Customers were informed a Product Support Specialist will contact them to make arrangements for a Field Service Technician to perform a test designed to identify if the table component affects the table mechanics. If the table is identified as having the affected component, the component will be replaced.
- Initiated
- 2014-07-10
- Root cause
- Component design/selection
- Terminated
- 2017-01-03
- Distribution
- Worldwide Distribution - US Nationwide in the states of CA, GA, FL, IL, IN, KS, KY, MD, MN, MO, MS, NC, NY, OH, TN, TX, VA and WI.
- Recall number
- Z-2209-2014
- Recall status
- Terminated
- CDRH record ID
- 128674
- FDA classified
- 2014-08-13
- Recall event number
- 68763