Philips Medical Systems North America Co. Phillips
FDA classification date
Aug 11, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification
Reason stated in the source
Monitor ceiling carriage may separate from the monitor ceiling suspension.
Source facts
Action
On 07/02/08, the firm sent a "Urgent - Device Correction" letter to all their consignees. The letter describes a product as X-Ray system with Monitor Ceiling Suspension and motorized height movement.
The consignees are informed that the fixation bolts supporting the monitor ceiling carriage may come loose creating a situation which may allow the monitor ceiling carriage to separate from motorized height Monitor Ceiling Suspension and that would cause serious injury to patient and user.
The firm has issued a mandatory Field Correction Action to correct the problem and the firm's technicians will inspect all ceiling suspension actuators for loose screws. If loose screws are found the actuator will be replaced.
Consignees should call Philips Call Center at 800-722-9377 #5, #4, #1, and Reference FCO72200126.
Initiated
2008-06-26
Root cause
Other
Terminated
2010-12-02
Distribution
United States and Canada.
Recall number
Z-2210-2010
Recall status
Terminated
CDRH record ID
72469
FDA classified
2010-08-11
Recall event number