- FDA identifier
- Z-2211-2009
- Company in FDA record
- Spacelabs Healthcare, Incorporated
- FDA classification date
- Sep 11, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The Spacelabs Medical Integrated ECG Cables, PN: 700-0008-22, 700-0008-23, 700-0008-24, and 700-0008-25 will not withstand the high voltage of a defibrillation shock.
Source facts
- Action
- Spacelabs Healthcare issued an "Urgent: Medical Device Recall" dated July 20, 2009. Consignees are being sent replacement ECG cables for each item ordered and are requested to replace the affected cables. Also, they are instructed to destroy and dispose of the recalled cables according to their own protocol or local regulations. For further information, contact Spacelabs Healthcare Technical Support at 1-800-522-7025, option 2 (US) or +86-512-87171888, extension 304 (outside US).
- Initiated
- 2009-07-23
- Root cause
- Process control
- Terminated
- 2011-07-21
- Distribution
- International Distribution -- Bangladesh, Turkey, Trinidad and Tobago, Romania, Guatemala, Algeria, Netherlands, Egypt, Chile, Venezuela, Spain, Colombia, India, the United Kingdom, Singapore, France and Ecuador.
- Recall number
- Z-2211-2009
- Recall status
- Terminated
- CDRH record ID
- 84707
- FDA classified
- 2009-09-11
- Recall event number
- 53124