Z-2211-2023: Contour next GEN Blood Glucose Monitoring System

Recall record Z-2211-2023 for Ascensia Diabetes Care US, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2211-2023
Company in FDA record
Ascensia Diabetes Care US, Inc.
FDA classification date
Jul 20, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Ascensia Diabetes Care has determined through customer complaints that 580 of 2900 meters from lot number DM01T033P are reconfigured from mg/dL to mmol/L. Among them, 579 meters distributed in the United States were packaged into the meter kit and had incorrect factory-set unit of measure where the meters display glucose results in mmol/L rather than mg/dL. The standard unit of measurement for the United States is mg/dL. If a consumer does not notice the incorrect unit of measurement, it is possible that the meters glucose measurement will be read as a lower blood glucose measurement than expected, and this may result in the patients glucose level remaining high, which can potentially lead to an injury requiring immediate medical intervention.

Source facts

Action
The firm sent recall notifications to customers dated 06/21/2023. Customers are asked to check their personal meters and entities are asked to check their stock to identify any affected devices. If you are aware of any patients or customers downstream who have received a recalled device make them aware of this recall. Anyone with a recalled unit can contact the firm at 1-800-348-8100 or by email at support@contournext.com to initiate return and replacement of devices. Response Forms were provided for customers to return via email to qualitysupport@ascensia.com.
Initiated
2023-05-19
Root cause
Process control
Distribution
Domestic: CA, CO, FL, IN, MO, NJ, NY, OH, RI, SC, TN, TX, & VA.
Recall number
Z-2211-2023
Recall status
Open, Classified
CDRH record ID
201321
FDA classified
2023-07-20
Recall event number
92536

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗