Skip to content

Public FDA record

Z-2212-2014: Device Brand Name: Distal Radius Fracture Repair System Device Classification Name: Plate, Fixation, Bon

Recall record Z-2212-2014 for Biomet, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2212-2014
Company in FDA record
Biomet, Inc.
FDA classification date
Aug 13, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm received complaints indicating that the drill guide becomes stuck in the jig. The drill guide can be placed in the jig about halfway before it becomes stuck.

Source facts

Action
On June 23, 2014 Biomet issued recall notification to their consignees via letter. Notification included affected product issue and instructions for return were provided. Distributors who had affected hospitals in their regions were notified and provided with a copy of the notice that was sent to the hospital. Consignees are being contacted on an individual basis in an attempt to reconcile the product 100%. A response form was provided to all consignees in order for Biomet to track returns. Non-responders will be contacted on an individual basis.
Initiated
2014-06-23
Root cause
Nonconforming Material/Component
Terminated
2015-07-14
Distribution
The product was distributed domestically and to internationally. Domestically to CA, PA, IA, CO, NY, MD, NC,AL, AZ, IL, GA, VA. Netherlands, Argentina,
Recall number
Z-2212-2014
Recall status
Terminated
CDRH record ID
128990
FDA classified
2014-08-13
Recall event number
68651

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗