Public FDA record

Z-2212-2026: Pediatric care bed; Product Designation: KayserBett IDA;

Recall record Z-2212-2026 for KAYSERBETTEN GMBH & CO. KG. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2212-2026
Company in FDA record
KAYSERBETTEN GMBH & CO. KG
FDA classification date
Jun 4, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

If the adjustment functions of the hand control, as instructed in the Manual, are not locked with the provided key when the patient (child) is unattended, there is an increased risk that the patient (child) or third parties (children) will operate the hand control and trap themselves or others under the bed frame or between the bed frame and the floor. This can lead to serious injuries or even death of the patient (child) or third parties (children).

Source facts

Action
On May 4, 2026 Urgent Field Safety Notice letters were sent to customers. Actions to be taken: Please inform all users and operators of a KayserBett IDA that by implementing the following safety measures, the above-mentioned risk to patients or third parties can be eliminated during further use of the product. If the patient is unattended, " the sleeping platform must be lowered to its lowest position. " the adjustment functions on the hand control must be locked using a key. " the doors must be closed. KayserBetten is implementing a corrective action to replace the hand control on all affected KayserBett IDA beds in the U.S. field with a modified hand control featuring an automatic locking function (Auto-Lock). With the new hand control, the adjustment functions lock automatically after each use, eliminating the dependence on the caregiver manually locking the hand control with a key. The hand control replacement will be performed by an authorized service technician at no cost to the customer. Your U.S. distributor, Mobility Unlimited Inc. (DBA KayserBetten US), will contact you to schedule the replacement. Please ensure in your organization that all users of the above-mentioned product and other persons to be informed are aware of this Field Safety Notice. If you have supplied the products to third parties, please forward a copy of this information or inform the contact person listed below. Please keep this information at least until the hand control replacement has been completed.
Initiated
2026-05-04
Root cause
Device Design
Distribution
Distribution US Nationwide and Canada.
Recall number
Z-2212-2026
Recall status
Open, Classified
CDRH record ID
219886
FDA classified
2026-06-04
Recall event number
98820

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗