Open original FDA record ↗Recall record
- FDA identifier
- Z-2213-2010
- Company in FDA record
- Minntech Corp
- FDA classification date
- Aug 12, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The DSD hook up Application Guide incorrectly recommends use of the Medivators DSD-110-HU0109 endoscope hookup to connect the Pentax EG-3630U Ultrasound Endoscope to a Medivators DSD endoscope reprocessor or Medivators Scope Buddy Endoscope Flushing Aid. This may result in inadequate disinfection during reprocessing, and subsequent infection risk to the patient. The recall was due to the misprinte
Source facts
- Action
- Consignees were sent a Medivators Reprocessing Systems "Urgent Medical Device Recall" letter date June 17, 2010. The letter was addressed to "Dear Medivators Customers". The letter described the problem and product involved. Recommended consignees to immediately examine the Pentax EG-3630U Ultrasound Endoscope hookups and inform all staff of the notice to ensure that the DSD-11--HU01109 is not used with the Pentax endoscope. Also requested consignees to complete and return the enclosed response form.
- Initiated
- 2010-06-15
- Root cause
- Labeling False and Misleading
- Terminated
- 2011-12-17
- Distribution
- AZ, AR, CA, CT, DE, FL, GA, ID, IL, IN, KS, KY, LA, MD, MA, MI, MN, MS, MO, NH, NJ, NY, NC, ND, OH, PA, SC, TN, TX, UT, VA, WA, WV, WI, WY.
- Recall number
- Z-2213-2010
- Recall status
- Terminated
- CDRH record ID
- 92897
- FDA classified
- 2010-08-12
- Recall event number
- 56189