Open original FDA record ↗Recall record
- FDA identifier
- Z-2215-2014
- Company in FDA record
- Zimmer, Inc.
- FDA classification date
- Aug 13, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Zimmer is initiating a voluntary recall of 20 production lots of Moore Hip Prosthesis Long Fenestrated Stems due to packaging design verification test failures. Specifically, 9 of 16 test samples failed simulated distribution and shipping testing when the Moore hip stem breached either the inner or outer package tray. Zimmer has not received any complaints of damaged package trays for these 20 lot
Source facts
- Action
- On July 1, 2014 Zimmer, Inc. issued recall notification to their consignees via letter. This notice was issued to affected distributors and hospitals with instructions for returning the affected product. Distributors were sent a letter identifying the issue and their responsibilities. These responsibilities include locating and removing the affected product in their territory, as well as identifying hospitals and surgeons who have previously implanted the affected product and providing related contact information (name, address, and phone number). Distributors are to return on-hand affected products to Zimmer and ensure all of their products are accounted for using the form provided in the letter. Hospital risk managers and surgeons were provided with a letter identifying the issue and their responsibilities. These responsibilities include locating the affected product and returning it to their Zimmer sales representative.
- Initiated
- 2014-07-01
- Root cause
- Packaging process control
- Terminated
- 2015-01-09
- Distribution
- Worldwide Distribution - US (nationwide) in the states of IN, WI, MN, OR, and TX, and one foreign consignee in Canada.
- Recall number
- Z-2215-2014
- Recall status
- Terminated
- CDRH record ID
- 128849
- FDA classified
- 2014-08-13
- Recall event number
- 68654