Z-2215-2024: 2nd Generation CentriMag Primary Console, REF: 201-90401, 201-90701, 201-90411, 201-90421, 201-30300, L20

Recall record Z-2215-2024 for Thoratec Switzerland GMBH. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2215-2024
Company in FDA record
Thoratec Switzerland GMBH
FDA classification date
Jun 27, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a component change, the console part of a circulatory support system may not conform to the IEC 61000-4-5 standard, so if subjected to a power surge above 1.8kV and up to 2.0 kV, the console may shut down completely without alarm, pumping would cease, and the patient would have to be switched to a backup circulatory support system.

Source facts

Action
On 5/22/24, correction notice distribution started. Customers were asked to do the following: If the console shuts down without alarms or alerts present, follow Instruction for Use (IFU) for Console and Motor malfunction as described in the Section Emergency / Troubleshooting , and switch to a backup console, which must be prepared as required backup components in the immediate vicinity. As stated in the IFU these backup systems should have a battery charge sufficient for at least one hour of operation, be connected to wall power, and be powered off. Note: When the backup console is put into use, the previously used console can be restarted and set up as a backup console. Per the IFU and current practice, a full console backup system is always required to be available in the immediate vicinity of each patient whenever the system is used. Firm is in the process of implementing an additional in-process test by Q3 of 2024, which will ensure that all systems which will be distributed are confirmed to withstand a surge up to 2.0 kV. Distribute this notice to those who need to be aware within your institution and to any organization where potentially affected devices may have been transferred in the event that devices were transferred elsewhere. Complete and return the acknowledgement form via email to CentriMagNotices@abbott.com Should you have any questions about this communication, please contact firm Customer Service at 1 (800) 456-1477
Initiated
2024-05-22
Root cause
Process control
Distribution
Worldwide - US Nationwide distribution included in the states of CA, PA, FL, IL, NY, AR, GA, WI, AZ, TN, TX, NC, MT, VA, MA, NJ, NE, WV, ME, OK, CO, OH, KC, WA, IN, CT, MI, OR, ID, MO, SC, KY, LA, NM, NV, VT, IA, MD and the countries of ALGERIA, AUSTRALIA, AUSTRIA, BANGLADESH, BULGARIA, CANADA, CAYMAN ISLANDS, CHILE, COLOMBIA, COSTA RICA, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, INDIA, IRAN, IRAQ, ISRAEL, ITALY, KENYA, KOREA, KUWAIT, LITHUANIA, MACEDONIA MALAYSIA, MARTINIQUE, MEXICO, MOROCCO, NETHERLANDS, NORWAY, OMAN, PERU, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TUNISIA, UNITED ARAB EMIRATES, UNITED KINGDOM.
Recall number
Z-2215-2024
Recall status
Open, Classified
CDRH record ID
207617
FDA classified
2024-06-27
Recall event number
94588

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗