Open original FDA record ↗Recall record
- FDA identifier
- Z-2216-2016
- Company in FDA record
- Roche Molecular Systems, Inc.
- FDA record creation date
- Jul 15, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Device was released with default pcal/gain settings for photometers instead of the pcal/gain settings that are required for release. The default settings can potentially lead to the generation of erroneous or invalid results.
Source facts
- Action
- Customers were notified of the recall via phone and email sent on 5/9/2016. They were instructed to take the following actions: Immediately discontinue use of the affected instruments until they have been reworked by a Roche Field Engineering Specialist. A Roche representative will contact you to schedule this service. Complete the attached fax form and fax it to 1-317-521-4815. File this Urgent Medical Device Correction (UMDC) for future reference. Please contact the Roche Support Network Customer Support Center 1-800-800-5973 if you have questions.
- Initiated
- 2016-05-09
- Root cause
- Under Investigation by firm
- Terminated
- 2021-03-24
- Distribution
- Distributed in NC and WA.
- Recall number
- Z-2216-2016
- Recall status
- Terminated
- CDRH record ID
- 146266
- FDA record created
- 2016-07-15
- Recall event number
- 74094