Open original FDA record ↗Recall record
- FDA identifier
- Z-2216-2017
- Company in FDA record
- Zimmer Biomet, Inc.
- FDA record creation date
- Jun 1, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Zimmer Biomet is conducting a voluntary recall for a single lot of the Modular Hybrid Glenoid Base Large for the Comprehensive Shoulder System after it was determined that two pieces from the lot were moved to finished goods and shipped despite having an open Non-Conformance Report (NCR). The NCR indicated that one piece from the lot had scratches and the other non-confirming threads.
Source facts
- Action
- Zimmer Biomet initiated a voluntary recall on April 26, 2017, for a single lot of the Modular Hybrid Glenoid Base Large for the Comprehensive Shoulder System after it was determined that two pieces from the lot were moved to finished goods and shipped despite having an open Non-Conformance Report (NCR). The NCR indicated that one piece from the lot had scratches and the other non-conforming threads. Notices were distributed via FedEx to customers. Risk Managers were instructed as follows: 1. Review the notification and ensure affected personnel are aware of the contents. 2. Assist your Zimmer Biomet sales representative quarantine all affected product. 3. Your Zimmer Biomet sales representative will remove the affected product from your facility. 4. Complete Attachment 1 Certificate of Acknowledgement. a. Return a digital copy within three (3) to corporatequality.postmarket@zimmerbiomet.com. b. Retain a copy of the Acknowledgement Form with your field action records in the event of a compliance audit of your facilities documentation. 5. If after reviewing this notice you have further questions or concerns please call 411 Technical Services at 574- 371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of the call center operating hours will receive a prompt to record a voicemail or be transferred to an on-call representative in the case of an emergency. Alternatively, your questions may be sent by email to corporatequality.postmarket@zimmerbiomet.com. Surgeons were instructed as follows: Review the notification for awareness of the contents. 2. There are no specific patient monitoring instructions related to this field action that are recommended beyond your existing follow up schedule. 3. Complete Attachment 1 Certificate of Acknowledgement. a. Return a digital copy to corporatequality.postmarket@zimmerbiomet.com. b. Retain a copy of the Acknowledgement Form with your field action records in the event of a co
- Initiated
- 2017-04-26
- Root cause
- Under Investigation by firm
- Terminated
- 2018-03-30
- Distribution
- Worldwide Distribution - US including OH AZ ID PA AR MN IL CA MD CA MN NY Foreign: Canada
- Recall number
- Z-2216-2017
- Recall status
- Terminated
- CDRH record ID
- 155038
- FDA record created
- 2017-06-01
- Recall event number
- 77119