Z-2217-2024: NovaGuide 2 Traditional Ultrasound , REF: NSC-TCDNG2

Recall record Z-2217-2024 for NovaSignal Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2217-2024
Company in FDA record
NovaSignal Corp.
FDA classification date
Jun 27, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to inadvertently disconnecting grounding cable of ultrasound system could result in electrical shock, electrical interference and/or equipment malfunction.

Source facts

Action
On May 6, 2024, NeuraSignal Inc. dba NovaSignal issued a "Urgent-Medical Device Recall" notification to affected consignees. Nova Signal asked consignees to take the following actions: 1. The NeuraSignal service and customer support team will schedule a time for the onsite field correction. 2. In the timeframe leading up to the scheduled onsite field correction, the system can be used as intended. 2a. Users of the device should refrain from accessing the enclosed cable connections during routine operation. This may lead to unintentional disconnection of the chassis ground cable from the AC mains power distribution unit. It is recommended that the cart compartment remains locked while the device is commissioned for use. 3. The site should confirm that the grounding cable has not been unintentionally disconnected and maintains a secure connection to the power distribution unit 4. If necessary, resources from the Biomedical Equipment department may perform re-qualification or pre-installation qualification testing, to ensure the system is adequately grounded.
Initiated
2024-05-06
Root cause
Device Design
Distribution
Worldwide - US Nationwide distribution in the states of MA, TX, NY, MN, AZ, AL, CO, CA, OH, MO, FL, TN, NJ, GA, OK, KY, WA and the country of Canada.
Recall number
Z-2217-2024
Recall status
Open, Classified
CDRH record ID
207962
FDA classified
2024-06-27
Recall event number
94673

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗