Open original FDA record ↗Recall record
- FDA identifier
- Z-2218-2010
- Company in FDA record
- AngioScore Inc.
- FDA classification date
- Aug 18, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Separations of the PTCA catheter shaft at the guide wire exit port have occurred during treatment of coronary artery stenosis. Fragments of the catheter may become lodged in coronary arteries, which resulting in serious patient outcome and require emergency coronary bypass surgery and medical intervention to remove the retained device fragments.
Source facts
- Action
- AngioScore sent an "Urgent: Medical Device Voluntary Recall" dated December 4, 2009 addressed to "Catheterization Lab Manager" at consignees. The Recall notification letter instructed consignees to inspect their inventory to detect product from the involved lot, and segregate it from their inventory. Consignees were asked to record all affected product and fax return forms to the firm. AngioScore Customer Service will contact consignees and provide instructions for return and replacement of affected product.. Consignees can contact AngioScore at 877-264-4692.
- Initiated
- 2009-12-04
- Root cause
- Process design
- Terminated
- 2011-06-21
- Distribution
- Nationwide distribution.
- Recall number
- Z-2218-2010
- Recall status
- Terminated
- CDRH record ID
- 89016
- FDA classified
- 2010-08-18
- Recall event number
- 54687