Open original FDA record ↗Recall record
- FDA identifier
- Z-2218-2011
- Company in FDA record
- International Technidyne Corp.
- FDA classification date
- May 19, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
An incorrect calibration code that affects the total hemoglobin measurement was applied to affected Avoximeter instruments.
Source facts
- Action
- International Technidyne Corp. sent a letter dated March 23, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to check their inventory to determine if they have any of the affected product. If they have any of the affected product, customers were instructed to stop using it and remove it from their inventory. Customers were also instructed to complete the attached Customer Account Tracking Form and return it by fax to (732-635-0144) or e-mail (techsupport@itcmed.com) or mail to: ITC Technical Support 20 Corporate Place South Piscataway, New Jersey 08854 For any questions call the Distributor or ITC Nexus Dx Technical Support at 800-631-5945 (US) or 732-548-5700 (International), ext 4707.
- Initiated
- 2011-03-23
- Root cause
- Device Design
- Terminated
- 2012-10-25
- Distribution
- Worldwide Distribution - USA including TX, KS, OH, OR, FL, GA, NY, MA, and DC and the countries of Taiwan and Thailand
- Recall number
- Z-2218-2011
- Recall status
- Terminated
- CDRH record ID
- 99574
- FDA classified
- 2011-05-19
- Recall event number
- 58549