Z-2218-2024: Prelude Roadster Guide Sheath REF: -PG4F45S -PG4F90S -PG5F45MP -PG5F45R -PG5F45S -PG5F45S018 -PG5F90MP -P

Recall record Z-2218-2024 for Merit Medical Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2218-2024
Company in FDA record
Merit Medical Systems, Inc.
FDA classification date
Jun 27, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to older version of Instruction for Use (IFU) instructing to ensure that the value and the sheath are tightly connected before use. This may cause users to over-tightening the luer which may result in the swivel nut detaching from the HVA body.

Source facts

Action
On 05/22/2024 and on 05/28/2024, the firm emailed an "URGENT MEDICAL DEVICE RECALL NOTICE" to customers informing them that the Instructions for Use (IFU) instructs users to "ensure that valve and the sheath are tightly connected before use" which may cause over-tightening of the luer fitting that may result in the swivel nut detaching from the Hemostasis valve adapter. Customers are instructed to: 1. Read and understand this letter and the IFU provided. There is no need to return product to Merit. 2. Post this notice and updated IFU on or near the affected products so users are aware of the recall. 3. Ensure that applicable personnel within your organization are made aware of this communication. 4. If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice and updated IFU is immediately shared with them. Additional distribution details may be required by health authorities. 5. If you require paper IFUs, please contact RESPONSE@merit.com. 6. Please fill out, scan and email the completed Customer Response Form to Customer Service at RESPONSE@merit.com within 7 calendar days. If any questions - contact the Merit Sales Representative or Merit Customer Service via email at RESPONSE@merit.com or via phone at 801-208-4381 Hours: 6 am to 6 pm MST/Mon -Fri.
Initiated
2024-05-22
Root cause
Device Design
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT. NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, and WV. The countries of Argentina, Brazil, Canada, Cayman Island, Chile, Colombia, Costa Rica, Dominican Republic, Ecuador, Hong Kong, Republic of Korea, Mexico, New Zealand, Peru, South Africa, and Uruguay.
Recall number
Z-2218-2024
Recall status
Open, Classified
CDRH record ID
207999
FDA classified
2024-06-27
Recall event number
94684

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗