An issue was identified with the TomoTherapy HI-Art System. In the event a patient or DQA plan has a moved image, roll adjustments applied during registration will be incorrect.
The Planning Station Plan Settings and DQA Setup tabs allow for images to be moved during planning. During registration when roll is applied on moved images, the Operator Station incorrectly rolls the image about
Source facts
Action
Consignee was sent a TomoTherapy "Urgent Medical device Correction Field Safety Notice' dated December 01, 2009. The letter described the Issue, Product Affected, Recommended Action and Resolution.
Initiated
2009-12-01
Root cause
Software design
Terminated
2011-12-17
Distribution
VA
Recall number
Z-2219-2010
Recall status
Terminated
CDRH record ID
92786
FDA classified
2010-08-12
Recall event number
56166