Philips Medical Systems North America Co. Phillips
FDA classification date
May 19, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification
Reason stated in the source
The PercuNav Field Generator Stand could tip over during set up, during use, or during transport.
Source facts
Action
On 4/4/11, Philips Healthcare began sending out the URGENT-Medical Device Correction notification letter to their consignees.
Philips informed their customers of a potential tipping hazard of the PercuNav Field Generator Stand.
The notification letter also provides instructions to customers to ensure the articulating Mounting Arm is positioned so the Field Generator is not further than the specified distance from the back of the clamp on the stand's central pole during set up or during use.
Customers are instructed to apply a provided label on stand to warn of a potential for tipping due to excessive force. For transport, customers are instructed to remove the Field Generator from the Mounting Arm and place the Field Generator in the storage compartment on the system cart. The Mounting Arm should then be collapsed and folded into a 90 degree angle.
Customers are informed that a Philips field service representative will contact them to schedule installation of a service kit on the stand, free of charge, once service kits are available.
Customers can call Philips Customer Service at 1-866-767-7822 and reference to FCO 79500197 for any questions about this recall.
Initiated
2011-04-01
Root cause
Device Design
Terminated
2012-03-15
Distribution
Worldwide Distribution -- USA, Australia, Austria, Belgium, China, France, Germany, Hong Kong, India, Korea, Netherlands, Singapore, Sweden, Switzerland, Taiwan, Thailand, and United Kingdom.
Recall number
Z-2219-2011
Recall status
Terminated
CDRH record ID
99329
FDA classified
2011-05-19
Recall event number
58429
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.