Open original FDA record ↗Recall record
- FDA identifier
- Z-2219-2012
- Company in FDA record
- Siemens Medical Solutions, USA, Inc
- FDA classification date
- Aug 17, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
If multiple uncompressed images/clips are sent to the syngo Dynamics server, around the same time, it is possible for an image from a patient to be placed in another patient's study.
Source facts
- Action
- On 07/13/2012, Siemens issued a "Customer Safety Advisory Notice" to affected consignees according to Update Instruction SY062/121S. The notice informs the customers of the potential issue with the identified affected product and provides details regarding the resolution. A software update to relieve the symptom of the issue will be distributed and installed by a Siemens Service Engineer on all affected systems under Update Instruction SY063/121S. A software update to correct this issue will be distributed and installed by a Siemens Service Engineer on all affected systems under Update Instruction SY064/121S.
- Initiated
- 2012-07-13
- Root cause
- Software design
- Terminated
- 2013-07-15
- Distribution
- Nationwide Distribution -- MA, GA, MO, IA, CA, PA, OR, CO, NY, NJ, WA, UT, OH, WI, MI, KY, and NC.
- Recall number
- Z-2219-2012
- Recall status
- Terminated
- CDRH record ID
- 111087
- FDA classified
- 2012-08-17
- Recall event number
- 62581