Skip to content

Public FDA record

Z-2219-2024: Stryker Hoffman LRF Wire Tensioner REF 4933-9-100, a component of the Hoffmann LRF System.

Recall record Z-2219-2024 for Stryker GmbH. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2219-2024
Company in FDA record
Stryker GmbH
FDA classification date
Jun 27, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The adjustment ring on the device may become loose resulting in the wire tension not being able to be set correctly.

Source facts

Action
The recalling firm issued letters dated 5/9/2024 on 5/9/2024 and 5/11/2024 via email to sales reps and distributors and via FedEx ground to hospitals. The letter described the product, the issue, and potential risks. The consignees instructions were: (1) Immediately check your inventory to locate the affected product listed on the Business Reply Form, remove the product from its point of use, and isolate it to prevent accidental use; (2) Sign and return the enclosed Business Reply Form by email by 5/15/2024 to confirm receipt of the notification/documenting product disposition: (3) Upon receipt of the completed Business Reply Form, Stryker will contact the consignee to arrange for return of the product; (4) Maintain awareness of the communication internally until all required actions have been completed within the consignee facility; (5) If the consignee has further distributed the affected product, notify the applicable parties at once about the recall. If possible, inform Stryker if any of the subject devices have been distributed to other organizations, including contact details; and if the consignee is a distributor, note that the distributor is responsible for notifying their affected customers.
Initiated
2024-05-09
Root cause
Process change control
Distribution
Worldwide - US Nationwide distribution in the states of AZ, CA, CO, FL, GA, KY, MA, MI, NE, NJ, OH, OR, and TX. The countries of Germany, Japan, Netherlands, and United Kingdom.
Recall number
Z-2219-2024
Recall status
Open, Classified
CDRH record ID
208107
FDA classified
2024-06-27
Recall event number
94721

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗