Z-2220-2017: Large Stealth Chamber, Catalog Numbers: DS50-001-XL-T, DS50-002-XL-T, DS50-003-XL-T, DS50-102-XL-T, D

Recall record Z-2220-2017 for Iba Dosimetry. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2220-2017
Company in FDA record
Iba Dosimetry
FDA record creation date
Jun 2, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When the Stealth Chamber is used as a reference detector in relative dosimetry, deviation in the measured PDDs at shallow depths and geometric shadowing of measured profiles may appear.

Source facts

Action
IBA Dosimetry sent an Field Safety Notice dated April 21, 2017. The letter informed the consignee of the issue and provided the following instruction: If a Stealth Chamber was used during commissioning of Linear Accelerator and Treatment Planning System, the consignee was asked to consider referenced deviations for larger field sizes (>20 x 20 cm) and higher energies (>60MV). They were also instructed to recheck the influence of the PDD deviation and insufficient clearance for 40 x 40 profile on their beam model and its clinical relevancy. For further questions please call (901) 268-6430. Contact information Should you have questions or require additional information, please contact the Customer Service Team at:: South and Middle America, USA, and Canada : IBA Dosimetry America Bartlett, TN, USA US Service@.iba-group .com Phone: +1 901 386-2242 Fax: +1 901 382-9453 Asia Pacific, A Australia and New Zealand: IBA China, Dosimetry Dept. Beijing, China Service APAC@.iba-group .com Phone: +86 10 8080 9107 Fax: + 86 10 80809298 All other countries: IBA Dosimetry GmbH, Service Dept. Schwarzenbruck , Germany service@.iba-group .com Phone: +49 9128 607-38 Fax: +49 9128 607-26
Initiated
2017-04-26
Root cause
Device Design
Terminated
2018-06-11
Distribution
Worldwide Distribution - USA (nationwide) and to the countries of : Argentina, Australia, Austria, Canada, Cyprus, France, Germany, Hong Kong, Ireland, Islamic Republic of, Israel, Italy, Japan, Kingdom of Saudi Arabia, Kuwait, New Zealand, People's Republic of China, Philippines, Russia, Slovakia, South Africa, South Korea, Spain, Sweden, Taiwan, R.O.C., and United Kingdom.
Recall number
Z-2220-2017
Recall status
Terminated
CDRH record ID
155649
FDA record created
2017-06-02
Recall event number
77301

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗