Skip to content

Public FDA record

Z-2220-2024: CDC, Influenza A/H5 Subtyping Kit, Model/Catalog Number: FluIVD03-11

Recall record Z-2220-2024 for Centers For Disease Control and Prevention. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2220-2024
Company in FDA record
Centers For Disease Control and Prevention
FDA classification date
Jun 27, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The H5b component may fail to amplify, resulting in an inconclusive result.

Source facts

Action
The CDC notified customers via letter on 04/26/2024. Customers were instructed to halt use of the H5b component when testing a specimen with the CDC Influenza A/H5 subtyping kit, lot 220307. Proceed with testing with the InfA, H5a, and RP targets and do not delay resulting. If a specimen is positive for InfA and H5a, that is considered a presumptive H5 positive and should be sent to CDC immediately. Please reach out to flusupport@cdc.gov for further shipping guidance. If a specimen is positive for InfA and negative for H5a and has not been tested with the Influenza A Subtyping kit, proceed to testing the specimen with the Influenza A Subtyping kit. If a specimen is negative for all viral targets but positive for RP, that specimen is considered negative for Influenza A, H5. Please reach out to flusupport@cdc.gov with any inconclusive results.
Initiated
2024-04-26
Root cause
Under Investigation by firm
Terminated
2025-02-05
Distribution
US Nationwide distribution.
Recall number
Z-2220-2024
Recall status
Terminated
CDRH record ID
208126
FDA classified
2024-06-27
Recall event number
94739

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗