The H5b component may fail to amplify, resulting in an inconclusive result.
Source facts
Action
The CDC notified customers via letter on 04/26/2024. Customers were instructed to halt use of the H5b component when testing a specimen with the CDC Influenza A/H5 subtyping kit, lot 220307. Proceed with testing with the InfA, H5a, and RP targets and do not delay resulting. If a specimen is positive for InfA and H5a, that is considered a presumptive H5 positive and should be sent to CDC immediately. Please reach out to flusupport@cdc.gov for further shipping guidance. If a specimen is positive for InfA and negative for H5a and has not been tested with the Influenza A Subtyping kit, proceed to testing the specimen with the Influenza A Subtyping kit. If a specimen is negative for all viral targets but positive for RP, that specimen is considered negative for Influenza A, H5. Please reach out to flusupport@cdc.gov with any inconclusive results.
Initiated
2024-04-26
Root cause
Under Investigation by firm
Terminated
2025-02-05
Distribution
US Nationwide distribution.
Recall number
Z-2220-2024
Recall status
Terminated
CDRH record ID
208126
FDA classified
2024-06-27
Recall event number
94739