When an ARCHITECT c16000 System is connected to an ACCELERATOR APS System, the software on the ACCELERATOR APS System may not generate an error warning when a Sample Presentation/Sample Queue error occurs, which could result in contamination of subsequent samples.
Source facts
Action
Abbott Laboratories notified Customers of the affected device via a "Product Correction" letter dated October 27, 2008. Consignees provided with a temporary work around until a service representative upgrades the system software.
For further information, contact your local Abbott Laboratories, Inc. service representative.
Initiated
2008-10-27
Root cause
Software design
Terminated
2009-12-02
Distribution
International Distribution -- Spain, Italy, the United Kingdom, Sweden, Switzerland, the Netherlands, France, Canada and Scotland.
Recall number
Z-2221-2009
Recall status
Terminated
CDRH record ID
75490
FDA classified
2009-09-18
Recall event number
50488