Storage temperature listed on the label was incorrect. Label stated 2-30C.
Actual storage temperature is 2-8C.
Source facts
Action
Neogen Corporation sent a Product Notification dated September 25, 2016, to all affected consignees via email. The recall notification included a description of the reason for the recall, affected product, and information for the correction to product. Consignees were informed that an internal corrective action has been initiated and actions have been taken and implemented to prevent this situation from occurring again. The label has been corrected and existing inventory has been relabeled.
For questions regarding this recall call 800-234-5333, ext 2362.
Initiated
2016-09-25
Root cause
Error in labeling
Terminated
2017-06-02
Distribution
Worldwide Distribution - US to AZ; Internationally to ECUADOR; CANADA; UK; BRAZIL
Recall number
Z-2221-2017
Recall status
Terminated
CDRH record ID
155596
FDA record created
2017-06-02
Recall event number
77276
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.