Skip to content

Public FDA record

Z-2221-2019: Regard custom surgical pack labeled as "LAPAROSCOPIC - OSU" including recalled Stryker part numbers 0620

Recall record Z-2221-2019 for ROi Consolidated Service Center (CSC). View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2221-2019
Company in FDA record
ROi Consolidated Service Center (CSC)
FDA record creation date
Aug 8, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Resource Optimization & Innovation (ROi) has received a recall notice from Stryker regarding tube sets potentially detaching from the cassette, which can lead to leaking gas or a hissing sound.

Source facts

Action
ROi, Consolidated Service Center ROi, notified customers on about 03/21/2019. Each email included a copy of the ROi "MEDICAL DEVCIE RECALL NOTIFICATION" dated March 21, 2019 as well as the Stryker recall notification dated March 7, 2019. ROi notification letter instructed customers to review inventory to determine if the affected ROi Custom Pack item and lot numbers are in inventory, quarantine all affected packs on hand, and notify ROi of the number of each affected lot on hand by completing and returning "Attachment A" via email to lacy.stewart@roiscs.com. Supplemental labels for affected packs will be issued, which will instruct the end user to remove the affected Stryker component(s) and to use single sterile replacement product form inventory. If there is not any affected inventory on hand, customers are requested to still return Attachment A indicating "0" affected inventory at their facility.
Initiated
2019-03-21
Root cause
Nonconforming Material/Component
Distribution
Distributed to LA and MO
Recall number
Z-2221-2019
Recall status
Open, Classified
CDRH record ID
171726
FDA record created
2019-08-08
Recall event number
82499

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗