Fluid may enter the multi-functional handle (Opti Grip) when using the Axiom Luminos TF or Sireskop SD system, causing unintended reaction or movement, increased radiation or useless images.
Source facts
Action
Siemens Healthcare Sector sent an "Urgent Field Safety Notice" to consignees describing the issue and informing them to take specific precautions when performing interventional procedures to avoid fluids penetrating the system. The multi-functional handle (Opti Grip) will be replaced through implementation of Update Instructions AX030/10/S. The modified version of the handle resists fluids from entering the handle and prevents malfunction.
Initiated
2010-07-13
Root cause
Device Design
Terminated
2012-09-29
Distribution
Nationwide distribution: all 50 US states and PR.
Recall number
Z-2222-2010
Recall status
Terminated
CDRH record ID
93079
FDA classified
2010-08-13
Recall event number
56256