Z-2222-2019: SmartRelease Endoscopic Carpal Tunnel Release Handpiece (ECTR handpiece), part numbers 81014 and 83014

Recall record Z-2222-2019 for MicroAire Surgical Instruments, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2222-2019
Company in FDA record
MicroAire Surgical Instruments, LLC
FDA record creation date
Aug 8, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There was a higher than normal incidence rate of the blade not retracting when tightening.

Source facts

Action
Urgent Medical Device Recall notification letters were sent to customers on 6/26/19. Additional customers identified later sent letters on 7/17/19. Regardless of the above IFU information, to be on the safe side, MicroAire is performing this recall and asking that you please complete and send back page two of this letter to notification@microaire.com or via fax at 800-648-4309. We will rework the handpiece (by replacing the handpiece head) and return it to you. If you have any questions about this Recall, please contact MicroAire Customer Service at 1-800-722-0822 between the hours of 8:30 AM and 5:00 PM (EST) Monday through Friday.
Initiated
2019-06-26
Root cause
Component change control
Terminated
2020-04-29
Distribution
The products were distributed to the following US states: CA, CT, FL, GA, IA, IL, IN, MD, MI, MN, MS, MT, NC, NJ, NY, OH, OR, PA, SC, TN, TX, UT, VA, WA, WI, and WY. The products were distributed to the following foreign countries: Australia, Czech Republic, France, Germany, and Taiwan.
Recall number
Z-2222-2019
Recall status
Terminated
CDRH record ID
174547
FDA record created
2019-08-08
Recall event number
83283

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗