Reason stated in the source
a gap between the conveyor belt and the rack tray table (plastic part) may cause vibration of the 5 position racks during the transport. This issue can pose a potential risk to involved operators and can also lead to potential cross contamination of samples in the affected racks.
Source facts
- Action
- Roche sent an Urgent Medical Device Correction letter dated May 9, 2017, to all affected consignees via e-mail. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Consignees were asked to complete the attached form and fax it to 877-906-8982 or email to Indianapolis.cd_tsc@roche.com. Consignees with questions were instructed to contact the Roche Support Network Customer Support Center at 1-800-428-2336.
- Initiated
- 2017-05-09
- Root cause
- Device Design
- Terminated
- 2018-02-02
- Distribution
- US distribution to WA only.