Open original FDA record ↗Recall record
- FDA identifier
- Z-2223-2024
- Company in FDA record
- Cook Incorporated
- FDA classification date
- Jun 27, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Product labels state the incorrect expiration dates that extends beyond their true expiration date
Source facts
- Action
- Cook issued Urgent Medical Device Correction letter mailed via courier to the impacted U.S consignee on 5/16/24. Letter states reason for recall, health risk and action to take: Actions to be Taken by the Customer 1. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected products have been transferred. 2. Examine inventory immediately to determine if you have affected product(s) in your inventory that remain unused. If you have affected product(s) remaining within your inventory: a. You should retain a copy of this letter in an easily accessible location and/or attach a copy of this letter to the affected product. b. You may continue to use affected products. Cook Medical is not requesting product be returned. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter regardless of whether you have affected product(s) on hand. You may return the Form via fax (812.339.7316) or email (FieldActionsNA@CookMedical.com).
- Initiated
- 2024-05-15
- Root cause
- Labeling Change Control
- Distribution
- MA Foreign: Spain, and Netherlands
- Recall number
- Z-2223-2024
- Recall status
- Open, Classified
- CDRH record ID
- 208025
- FDA classified
- 2024-06-27
- Recall event number
- 94696