Skip to content

Public FDA record

Z-2223-2026: LVP software of the Ivenix Infusion System (IIS); Product Code LVP-SW-0005.

Recall record Z-2223-2026 for Fresenius Kabi USA, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2223-2026
Company in FDA record
Fresenius Kabi USA, LLC
FDA classification date
Jun 3, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Potential software anomaly that can cause the pump to report a false battery-health value (typically 69%) and trigger a battery-depletion alarm that will shut down the pump unless it is reconnected to AC mains power.

Source facts

Action
Consignees were notified on about 05/06/2026 via letter. Consignees were informed of the issue, risk associated, and instructed to keep the Ivenix pump plugged in at all times rather than rely on battery and if a "Battery Extremely Low" alarm occurs while on battery, immediately connect the pump to AC. For any transport of a patient on an affected pump, have a plan for AC access. For patients receiving critical or short half-life infusion therapies, ensure an RN accompanies the patient during transport. Consignees were also instructed to notify other users if applicable, notify customers if affected units were transferred or distributed, and post the provided Safety Tip Sheet, Battery Health Software Anomaly in the Ivenix Large Volume Pump (LVP) notice alongside this letter in all areas where affected pumps are stored and used. consignees were asked to complete and return the provided Customer Reply Form.
Initiated
2026-05-06
Root cause
Software design (manufacturing process)
Distribution
US Nationwide distribution in the states of CA, CO, FL, GA, GA, ID, IL, MD, MI, MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI.
Recall number
Z-2223-2026
Recall status
Open, Classified
CDRH record ID
220229
FDA classified
2026-06-03
Recall event number
98929

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗