Skip to content

Public FDA record

Z-2224-2017: Venture RX Catheter, Model 5820; Venture OTW Catheter, Model 5821; Venture CS Catheter, Model 5822. Ster

Recall record Z-2224-2017 for Vascular Solutions, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2224-2017
Company in FDA record
Vascular Solutions, Inc.
FDA classification date
May 22, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Vascular Solutions, Inc. (VSI) is voluntarily removing all lots of Venture catheters due to a potential problem. Venture RX (Model 5820), Venture OTW (Model 5821), and Venture CS (Model 5822) catheters. After an internal investigation, VSI determined that material used in catheter construction may detach from within the lumen of the distal tip during use.

Source facts

Action
The firm, Vascular Solutions, sent an "URGENT: Medical Device Recall" letter dated April 25, 2017, to all affected consignees. The letter described the problem and the product involved in the recall and actions to be taken. Consignees were instructed to identify location of all unused product in your possession; immediately cease further use or distribution, remove any unused product , place it in a secure area, and complete and return the VSI Account Inventory Form by email to: customerservice@vascom or fax to (763)656-4251. VSI Customer Service Department will contact consignees for return authorization of the product. If you have any specific questions or concerns regarding this matter, please contact your local sales representative or call 763-656-4300.
Initiated
2017-04-25
Root cause
Process control
Terminated
2019-11-05
Distribution
Worldwide Distribution - US (nationwide): including states of: AK, AL, AR, AZ, CA, CO, CT, DC,FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MN, MO, ME, MI, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY; and countries of: AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CANADA, CZECH REPUBLIC, EGYPT, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, IRELAND, ISRAEL, ITALY, LEBANON, NETHERLANDS, NORWAY, POLAND, RUSSIA, SAUDI ARABIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TURKEY,and UK.
Recall number
Z-2224-2017
Recall status
Terminated
CDRH record ID
155426
FDA classified
2017-05-22
Recall event number
77202

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗