Open original FDA record ↗Recall record
- FDA identifier
- Z-2224-2021
- Company in FDA record
- Integra LifeSciences Corp.
- FDA record creation date
- Aug 6, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Packaging-pouches with wrinkles across the Tyvek side may compromise the integrity of the pouch's sterile barrier
Source facts
- Action
- Integra notified consignees (Hospitals, Government facilities, and Distributors) by letter via FedX 2 Day Mail on June 1, 2021 and email to Distributor. Letter states reason for recall, health risk and action to take: We request that you take the following actions: 1. Customers a. If you do have units of the affected product listed in Table 1, remove it immediately from service. b. If you do have affected product, check the box on the enclosed form I do have affected product. Record the total quantity of the affected product that you have. c. If you do not have affected product, check the box, I do not have affected product. d. Complete the attached Acknowledgement Form and return to FCA2@integralife.com or FAX to 1-609-750-4220. e. Keep a copy of the form for your records. f. When your form is received, and it is noted that you have affected product, Customer Service will contact you and provide a Return Material Authorization (RMA) number and directions to return the affected product. A replacement order will also be placed for the quantity noted on the form. 2. Distributors a. If you do have the affected product listed in Table 1, remove the product from further distribution. b. Complete the attached form. If you do not have affected product, check the box, I do not have affected product. c. If you do have affected product, check the box I do have affected product. Record the total quantity of affected product you have. d. Complete the attached Acknowledgement Form and return to FCA2@integralife.com or FAX to 1-609-750-4220. e. Keep a copy of the form for your records. f. Please check your customer traceability records for shipments of above catalog numbers. g. Forward a copy of the enclosed Field Safety Notice to any of your customers that have purchased the above catalog and lot number. h. When the form from you and your customer is received, and it is noted that you or your customer has affected product, Customer Service will co
- Initiated
- 2021-06-01
- Root cause
- Process change control
- Terminated
- 2023-01-27
- Distribution
- US Nationwide Distribution - Foreign Distribution: Hong Kong
- Recall number
- Z-2224-2021
- Recall status
- Terminated
- CDRH record ID
- 188368
- FDA record created
- 2021-08-06
- Recall event number
- 88312