Open original FDA record ↗Recall record
- FDA identifier
- Z-2225-2010
- Company in FDA record
- Fresenius Medical Care North America
- FDA classification date
- Aug 15, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Power cords may fail if the prongs crack or fail at the plug, and may become a fire/burn/shock hazard or delay/interrupt therapy.
Source facts
- Action
- An urgent device recall notification was sent to all consignees. Customers are asked to inspect cords for defects, and are refered to preventive maintenance requirements in the manual. Customers who identify the defective cords will be instructed to discontinue use as soon as possible without jeopardizing patient care. Customers are instructed to monitor cords regularly, and pay attention to machines that are frequently plugged or unplugged. Personnel are instructed to not use the cord to unplug the device from the wall (yanking the plug from the wall) and to refrain from rolling over the cord or plug.
- Initiated
- 2009-11-20
- Root cause
- Component design/selection
- Terminated
- 2011-02-07
- Distribution
- Product distributed to 5404 customers throughout the US and the world.
- Recall number
- Z-2225-2010
- Recall status
- Terminated
- CDRH record ID
- 88695
- FDA classified
- 2010-08-15
- Recall event number
- 54546