Software anomaly - Under certain conditions, unwanted alterations to results could be applied when using the BCI Connect scripting feature.
Source facts
Action
bioMerieux in North Carolina issued an 'URGENT PRODUCT CORRECTION NOTICE FSCA 5173-1 - MAY 2021" to their U.S. consignees on 6/22/2021 via email or overnight mail. The notice described the issue and listed the criteria that would result in the MYLA system being impacted.
If all of the criteria are confirmed on the MYLA system, then the system will be impacted by this FSCA and the consignee is instructed to stop using the user scripting rules that change the value in field a4 (Final Category) and update the MYLA application to version 4.8.2.
Initiated
2021-06-22
Root cause
Software design
Terminated
2023-02-06
Distribution
Distribution in the U.S. was nationwide. There was also military and government distribution.
Recall number
Z-2225-2021
Recall status
Terminated
CDRH record ID
187589
FDA record created
2021-08-06
Recall event number
87985