Z-2225-2026: Philips Azurion systems not configured with an optional auxiliary pan handle. Includes the below product

Recall record Z-2225-2026 for PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2225-2026
Company in FDA record
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
FDA classification date
May 21, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential safety issue with Philips Azurion systems, where longitudinal and transverse table movements may be impacted due to mechanical wear of the Float Tabletop control on the control module.

Source facts

Action
Philips Medical notified consignees on about 04/28/2026 via certified mailed letter titled "URGENT Medical Device Correction." Consignees were made aware of the issue and instructed to circulate the notification to all users of the system for awareness, that affected systems may continue to be used in accordance with their Instructions for Use (IFU), if the required position cannot be achieved using the Float Tabletop control for table movement, use alternative system movements (e.g., stand movement) where clinically feasible to adjust the relative position between the system and the patient, and avoid contact with sharp edges in the event that the Float Tabletop control is damaged. Consignees were also instructed to maintain the notification letter with the documentation of the system until Philips corrects the system, ensure the notification is located where it can be viewed, notify customers if system has been transferred, and complete and return the provided Response Form. Philips will install an auxiliary pan handle on all affected systems. A Philips representative will contact consignees to schedule the implementation of this correction. Philips expects to begin implementation of the correction by July 2026.
Initiated
2026-04-28
Root cause
Device Design
Distribution
Nationwide distribution. International distribution to Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Canada, Chile, C¿te D'Ivoire, El Salvador, Estonia, Greece, Honduras, Hong Kong, Iceland, Indonesia, Iraq, Japan, Kyrgyzstan, Malaysia, Malta, Moldova, Nepal, New Zealand, Philippines, Qatar, Singapore, South Africa, Sri Lanka, Syrian Arab Republic, Tanzania, Turkmenistan, Vietnam, Yemen
Recall number
Z-2225-2026
Recall status
Open, Classified
CDRH record ID
219872
FDA classified
2026-05-21
Recall event number
98812

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗