Open original FDA record ↗Recall record
- FDA identifier
- Z-2226-2009
- Company in FDA record
- Applied Medical Resources Corp
- FDA classification date
- Sep 18, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Applied Medical is conducting a coluntary recall of two lot numbers, 1084426 and 1085506, of the 15mm Kii non-threaded optical access system, model C0R36, due to a potential weakness in the cannula tip, possibly resulting in tip breakage.
Source facts
- Action
- Applied Medical Resources Corp. notified Customers via an overnight "Voluntary Recall Notification" Letter dated June 9, 2009. The letter informed customers of the reason for recall and asked that the listed product be returned. The recall letter provided customers with a contact and telephone number for product return questions (1-949-713-8652) and a separate contact for regulatory questions (1-949-713-8369).
- Initiated
- 2009-06-04
- Root cause
- Component design/selection
- Terminated
- 2011-07-14
- Distribution
- Class II Recall - Worldwide -- United States, Australia, New Zealand, France, Canada, Norway and Great Britain.
- Recall number
- Z-2226-2009
- Recall status
- Terminated
- CDRH record ID
- 83977
- FDA classified
- 2009-09-18
- Recall event number
- 52774