Skip to content

Public FDA record

Z-2226-2010: Fresenius Granuflo 1 Mixer Dissolution Unit - Tested, and Granuflo Dissolution Tank, Model Numbers G047-8

Recall record Z-2226-2010 for Fresenius Medical Care North America. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2226-2010
Company in FDA record
Fresenius Medical Care North America
FDA classification date
Aug 15, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Power cords may fail if the prongs crack or fail at the plug, and may become a fire/burn/shock hazard or delay/interrupt therapy.

Source facts

Action
An urgent device recall notification was sent to all consignees. Customers are asked to inspect cords for defects, and are refered to preventive maintenance requirements in the manual. Customers who identify the defective cords will be instructed to discontinue use as soon as possible without jeopardizing patient care. Customers are instructed to monitor cords regularly, and pay attention to machines that are frequently plugged or unplugged. Personnel are instructed to not use the cord to unplug the device from the wall (yanking the plug from the wall) and to refrain from rolling over the cord or plug.
Initiated
2009-11-20
Root cause
Component design/selection
Terminated
2011-02-07
Distribution
Product distributed to 5404 customers throughout the US and the world.
Recall number
Z-2226-2010
Recall status
Terminated
CDRH record ID
88700
FDA classified
2010-08-15
Recall event number
54546

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗