Z-2226-2021: ETEST IMIPENEM RELEBACTAM. in vitro diagnostic

Recall record Z-2226-2021 for Biomerieux Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2226-2021
Company in FDA record
Biomerieux Inc
FDA record creation date
Aug 6, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Major errors (Resistant result instead of Susceptible result) were observed with Pseudomonas aeruginosa strains.

Source facts

Action
The recall notification was issued to the impacted subsidiaries and distributors on 24JUN2021. The notification includes a Customer Letter to be translated into local language and sent to each impacted customer. The customer letter will inform the customers about the MIC overestimation and false resistant results issue and will be instructed to: - Perform a retest using an alternative method on P. aeruginosa strains for which they obtained resistant or intermediate results with ETEST IMIPENEM RELEBACTAM ref. 420925 and 420927 - Among tests previously performed, identify any possible false resistant/intermediate results that may have occurred, to analyze the related risks and to determine appropriate actions if relevant.
Initiated
2021-06-24
Root cause
Under Investigation by firm
Terminated
2025-02-10
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AR, AZ, CA, CO, DE, FL, GA, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, NC, NE, NH, NJ, NV, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, WI, WV and Hawaii. The countries of Austria, Belgium, Croatia, Czech Republic, France, Germany, Greece, Italy, Netherlands, Poland, Portugal, Spain, Sweden, Switzerland and United Kingdom.
Recall number
Z-2226-2021
Recall status
Terminated
CDRH record ID
188207
FDA record created
2021-08-06
Recall event number
88242

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗