Z-2226-2026: Silastic Foley Catheters Silastic Brand Foley Catheter 30cc REF: 33416 33418 33420 33422 33424 Silastic

Recall record Z-2226-2026 for C.R. Bard Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2226-2026
Company in FDA record
C.R. Bard Inc
FDA classification date
May 21, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to stain present on the surface of affected foley catheters.

Source facts

Action
On 04/23/2026, the firm sent via region courier services "URGENT Medical Device Recall" Letters informing customers that BD has received complaints, and confirmed through internal inspection that stains are present on the surface of the affected silastic catheters. The cause of the stains was determined to be an increased amount of a raw material used in the manufacturing process. This raw material may be harmful if present in increased amounts on the surface of the catheter. Customers are instructed to: 1. Immediately Discontinue Use 2. Please check all inventory locations within your institution for the Bard Silastic Foley Catheters listed in Attachment 2. 3. Immediately quarantine and destroy all devices within your facility s control per your facility s procedures. 4. Share this notice with any users of the product within your facilities to ensure they are also aware of this Medical Device Correction. 5. If you have purchased this product through a distributor, please contact this distributor to request appropriate credit. 6. Complete and return the attached Customer Response Form to the BD contact noted on the form confirming acknowledgement of the correction notification, whether or not you have any affected product, so that BD may confirm your receipt of this notification and provide credit. For Questions: Product Use Inquiries - 1-800-555-7422 option #2 available 24hours/7 days per week Product Complaints, Technical Support - 1-844-8BD-LIFE (1-844-823-5433) Say Product Complaint when prompted M-F 8am - 5pm CT or Email: productcomplaints@bd.com
Initiated
2026-04-23
Root cause
Material/Component Contamination
Distribution
U.S.: AK, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, and WI O.U.S.: Canada
Recall number
Z-2226-2026
Recall status
Open, Classified
CDRH record ID
219556
FDA classified
2026-05-21
Recall event number
98716

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗