Open original FDA record ↗Recall record
- FDA identifier
- Z-2227-2008
- Company in FDA record
- Datex Ohmeda, Inc. dba GE Healthcare
- FDA classification date
- Sep 11, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Medical device software may be subject to signal interference of the Hands Free Alarm Silence feature that could impact patient safety.
Source facts
- Action
- The recalling firm notified Hospital Administrators, NICU Nursing Manager/Labor & Delivery Nursing Managers and Heads of Biomedical Engineering departments at consignee locations by certified letter on 05/20/08. The notification was flagged as "Urgent Medical Device Correction" and advised consignees of safety issues and recommended actions when using the device. The notification advised that the safest way to mitigate the potential interference issue is to disable the Hands Free Alarm Silence. Users were informed that once a permanent correction action is identified, service representatives will implement the correction. For assistance, contact GE Healthcare at 1-800-345-2700.
- Initiated
- 2008-05-20
- Root cause
- Software design
- Terminated
- 2012-05-11
- Distribution
- The medical devices were distributed to hospitals nationwide.
- Recall number
- Z-2227-2008
- Recall status
- Terminated
- CDRH record ID
- 71706
- FDA classified
- 2008-09-11
- Recall event number
- 48603