Open original FDA record ↗Recall record
- FDA identifier
- Z-2227-2012
- Company in FDA record
- Beckman Coulter Inc.
- FDA classification date
- Aug 20, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class III
Reason stated in the source
A recall was initiated because Beckman Coulter has identified a negative bias in test results when comparing elevated Total T3 patient samples diluted with Sample Diluent A to samples diluted with Access Total T3 Calibrator S0.
Source facts
- Action
- Beckman Coulter sent a Product Correction letter dated October 11, 2011, to all affected customers. The PCA letter informed the customers of the problem identified and the actions to be taken. Customers were informed to stop SDA usage as an alternative dilutent and that the S0 dilutent is to be used for manual dilution of the patient sample for Total T3. They will also be informed to no longer use the On Board Dilution (OBD) option for Total T3. Customers were instructed to complete and return the enclosed response form within 10 days. Customers with questions regarding this notification were instructed to contact Customer Support Center at http://www.beckmancoulter.com/customersupport/support or call (800) 854-3633 in the US and Canada. Outside of the US and Canada, customers were instructed to contact their local Beckman Coulter Representative. For questions regarding this recall call 714-961-4941.
- Initiated
- 2011-10-11
- Root cause
- Device Design
- Terminated
- 2013-03-14
- Distribution
- Worldwide Distribution - USA (nationwide), Canada, and internationally.
- Recall number
- Z-2227-2012
- Recall status
- Terminated
- CDRH record ID
- 111723
- FDA classified
- 2012-08-20
- Recall event number
- 62791