Z-2227-2012: Access Total T3 Reagent, PN: 33830 The Access Total T3 assay is used for the quantitative determinatio

Recall record Z-2227-2012 for Beckman Coulter Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2227-2012
Company in FDA record
Beckman Coulter Inc.
FDA classification date
Aug 20, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class III

Reason stated in the source

A recall was initiated because Beckman Coulter has identified a negative bias in test results when comparing elevated Total T3 patient samples diluted with Sample Diluent A to samples diluted with Access Total T3 Calibrator S0.

Source facts

Action
Beckman Coulter sent a Product Correction letter dated October 11, 2011, to all affected customers. The PCA letter informed the customers of the problem identified and the actions to be taken. Customers were informed to stop SDA usage as an alternative dilutent and that the S0 dilutent is to be used for manual dilution of the patient sample for Total T3. They will also be informed to no longer use the On Board Dilution (OBD) option for Total T3. Customers were instructed to complete and return the enclosed response form within 10 days. Customers with questions regarding this notification were instructed to contact Customer Support Center at http://www.beckmancoulter.com/customersupport/support or call (800) 854-3633 in the US and Canada. Outside of the US and Canada, customers were instructed to contact their local Beckman Coulter Representative. For questions regarding this recall call 714-961-4941.
Initiated
2011-10-11
Root cause
Device Design
Terminated
2013-03-14
Distribution
Worldwide Distribution - USA (nationwide), Canada, and internationally.
Recall number
Z-2227-2012
Recall status
Terminated
CDRH record ID
111723
FDA classified
2012-08-20
Recall event number
62791

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗