Skip to content

Public FDA record

Z-2227-2017: Sesamoid Plasty, Stereotaxic Instrument The Sesamoid Plasty workstation is used to run the tracking softw

Recall record Z-2227-2017 for Orthosoft, Inc. dba Zimmer CAS. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2227-2017
Company in FDA record
Orthosoft, Inc. dba Zimmer CAS
FDA record creation date
Jun 2, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Zimmer CAS voluntarily conducted a retrospective recall of the Sesamoid Plasty due to the risk of the top of the workstation to potentially fall over during disassembly.

Source facts

Action
Zimmer CAS conducted a retrospective recall for the Sesamoid Plasty CAS workstation due to the possible risk of injury if the workstation is not properly disassembled. The firm mailed notices dated June 4, 2008. The original Sesamoid Plasty design did not include a safety design feature which is now available on current Sesamoid Plasty Workstations. The Safety design feature is a mechanism that prevents a latch lock at the base of the workstation column from being opened to allow folding of the workstation column for transport after use unless the camera and its arm are first removed from the top of the workstation. A safety label was initially used instead that instructed the user of the recommended disassembly sequence (i.e. camera and arm first). Once consignees receive their replacement system, they were instructed to contact Zimmer CAS Customer Service to assist them with the return (RMA) of the Sesamoid Plasty Workstation. For any questions or concerns please contact the Zimmer CAS customer service, 1-866-336-7846.
Initiated
2008-06-04
Root cause
Device Design
Terminated
2017-06-02
Distribution
Worldwide Distribution - US including FL, CA Foreign: Austrailia, New Zealand, France, Brazil
Recall number
Z-2227-2017
Recall status
Terminated
CDRH record ID
154865
FDA record created
2017-06-02
Recall event number
75609

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗